Quantity: Activity per unit area (C-14)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2001 (1 - 100) Bq.cm-2 3.6 %
  • C-14 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (Co-60)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2002 (10 - 100) Bq.cm-2 3.2 %
  • Co-60 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (Sr-90)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2003 (10 - 100) Bq.cm-2 3.2 %
  • Sr-90 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (Cs-137)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2004 (10 - 100) Bq.cm-2 3.2 %
  • Cs-137 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (Pm-147)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2005 (10 - 100) Bq.cm-2 3.2 %
  • Pm-147 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (Tl-204)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2006 (10 - 100) Bq.cm-2 3.6 %
  • Tl-204 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (Pu-239)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2007 (1 - 10) Bq.cm-2 3.6 %
  • Pu-239 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (Am-241)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2008 (1 - 10) Bq.cm-2 3.6 %
  • Am-241 rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Activity per unit area (U-nat)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2009 (1 - 5) Bq.cm-2 4 %
  • U-nat rectangular, 110 mm x 150 mm, or circular, diameter 23 mm to 45 mm
115-MP-C007
Quantity: Efficiency of contamination monitors (C-14)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2010 (0 - 1) s-1.Bq-1 10 %
  • C-14 detection beta radiation
115-MP-C007
Quantity: Efficiency of contamination monitors (Cs-1137)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2011 (0 - 1) s-1.Bq-1 10 %
  • Cs-137 detection beta radiation
115-MP-C007
Quantity: Efficiency of contamination monitors (Sr-90)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2012 (0 - 1) s-1.Bq-1 10 %
  • Sr-90 detection beta radiation
115-MP-C007
Quantity: Efficiency of contamination monitors (Am-241)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2013 (0 - 1) s-1.Bq-1 10 %
  • Am-241 detection alpha radiation
115-MP-C007
Quantity: Activity per unit mass (H-3)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2014 (5000 - 5.0E+7) Bq.g-1 4 %
  • H-3 tritiated water
115-MP-C007
Quantity: Activity per unit mass (Be-7)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2015 (5000 - 5.0E+7) Bq.g-1 2 %
  • Be-7 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (C-14)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2016 (5000 - 5.0E+7) Bq.g-1 3 %
  • C-14 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Na-22)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2017 (5000 - 5.0E+7) Bq.g-1 1.6 %
  • Na-22 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Na-24)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2018 (5000 - 5.0E+7) Bq.g-1 1.2 %
  • Na-24 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (P-32)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2019 (5000 - 5.0E+7) Bq.g-1 1.2 %
  • P-32 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (S-35)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2020 (5000 - 5.0E+7) Bq.g-1 2.6 %
  • S-35 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (K-42)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2021 (5000 - 5.0E+7) Bq.g-1 1.2 %
  • K-42 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Ca-45)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2022 (5000 - 5.0E+7) Bq.g-1 2.4 %
  • Ca-45 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Cr-51)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2023 (5000 - 5.0E+7) Bq.g-1 1.6 %
  • Cr-51 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Mn-54)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2024 (5000 - 5.0E+7) Bq.g-1 1.2 %
  • Mn-54 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Fe-55)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2025 (5000 - 5.0E+7) Bq.g-1 5.5 %
  • Fe-55 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Co-56)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2026 (5000 - 5.0E+7) Bq.g-1 4 %
  • Co-56 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Co-57)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2027 (5000 - 5.0E+7) Bq.g-1 1.6 %
  • Co-57 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Co-58)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2028 (5000 - 5.0E+7) Bq.g-1 2 %
  • Co-58 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Fe-59)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2029 (5000 - 5.0E+7) Bq.g-1 1.6 %
  • Fe-59 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Co-60)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2030 (5000 - 5.0E+7) Bq.g-1 0.8 %
  • Co-60 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Ni-63)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2031 (5000 - 5.0E+7) Bq.g-1 3 %
  • Ni-63 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Cu-64)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2032 (5000 - 5.0E+7) Bq.g-1 3.5 %
  • Cu-64 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Zn-65)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2033 (5000 - 5.0E+7) Bq.g-1 3 %
  • Zn-65 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Ga-67)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2034 (5000 - 5.0E+7) Bq.g-1 2.4 %
  • Ga-67 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Se-75)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2035 (5000 - 5.0E+7) Bq.g-1 2.5 %
  • Se-75 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (As-76)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2036 (5000 - 5.0E+7) Bq.g-1 2.6 %
  • As-76 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Br-82)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2037 (5000 - 5.0E+7) Bq.g-1 2 %
  • Br-82 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Sr-85)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2038 (5000 - 5.0E+7) Bq.g-1 2 %
  • Sr-85 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Rb-86)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2039 (5000 - 5.0E+7) Bq.g-1 2 %
  • Rb-86 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Y-88)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2040 (5000 - 5.0E+7) Bq.g-1 2.4 %
  • Y-88 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Sr-89)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2041 (5000 - 5.0E+7) Bq.g-1 1.2 %
  • Sr-89 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Sr-90)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2042 (5000 - 5.0E+7) Bq.g-1 1.2 %
  • Sr-90 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Y-90)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2043 (5000 - 5.0E+7) Bq.g-1 1.2 %
  • Y-90 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Zr-95)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2044 (5000 - 5.0E+7) Bq.g-1 2 %
  • Zr-95 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Nb-95)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2045 (5000 - 5.0E+7) Bq.g-1 1.2 %
  • Nb-95 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Mo-99)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2046 (5000 - 5.0E+7) Bq g-1 2 %
  • Mo-99 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Tc-99m)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2047 (5000 - 5.0E+7) Bq g-1 3 %
  • Tc-99m glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Ru-103)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2048 (5000 - 5.0E+7) Bq g-1 2.4 %
  • Ru-103 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Ru-106)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2049 (5000 - 5.0E+7) Bq g-1 2 %
  • Ru-106 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Cd-109)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2050 (5000 - 5.0E+7) Bq g-1 2.6 %
  • Cd-109 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (In-111)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2051 (5000 - 5.0E+7) Bq g-1 2.6 %
  • In-111 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Sn-113)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2052 (5000 - 5.0E+7) Bq g-1 2.4 %
  • Sn-113 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Sb-124)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2053 (5000 - 5.0E+7) Bq g-1 2.4 %
  • Sb-124 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (I-125)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2054 (5000 - 5.0E+7) Bq g-1 2 %
  • I-125 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (I-129)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2055 (5000 - 5.0E+7) Bq g-1 2 %
  • I-129 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (I-131)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2056 (5000 - 5.0E+7) Bq g-1 2 %
  • I-131 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Te-132)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2057 (5000 - 5.0E+7) Bq g-1 2.6 %
  • Te-132 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Ba-133)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2058 (5000 - 5.0E+7) Bq g-1 1.5 %
  • Ba-133 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Cs-134)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2059 (5000 - 5.0E+7) Bq g-1 1.6 %
  • Cs-134 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Cs-137)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2060 (5000 - 5.0E+7) Bq g-1 1.6 %
  • Cs-137 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Ce-139)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2061 (5000 - 5.0E+7) Bq g-1 1.6 %
  • Ce-139 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Ce-141)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2062 (5000 - 5.0E+7) Bq g-1 1.6 %
  • Ce-141 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Ce-144)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2063 (5000 - 5.0E+7) Bq g-1 2 %
  • Ce-144 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Pm-147)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2064 (5000 - 5.0E+7) Bq g-1 3 %
  • Pm-147 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Eu-152)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2065 (5000 - 5.0E+7) Bq g-1 1.6 %
  • Eu-152 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Gd-153)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2066 (5000 - 5.0E+7) Bq g-1 1.6 %
  • Gd-153 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Tm-170)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2067 (5000 - 5.0E+7) Bq g-1 1.6 %
  • Tm-170 glass ampoule
115-MP-C007
Quantity: Activity per unit mass (Ir-192)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2068 (5000 - 5.0E+7) Bq g-1 1.6 %
  • Ir-192 glass ampoule
115-MP-C007
Quantity: Hmotnostní aktivita (Hg-197)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2069 (5000 - 5.0E+7) Bq g-1 3 %
  • Hg-197 glass ampoule
115-MP-C007
Quantity: Hmotnostní aktivita (Hg-203)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2070 (5000 - 5.0E+7) Bq g-1 2 %
  • Hg-203 glass ampoule
115-MP-C007
Quantity: Hmotnostní aktivita (Tl-204)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2071 (5000 - 5.0E+7) Bq g-1 2.6 %
  • Tl-204 glass ampoule
115-MP-C007
Quantity: Hmotnostní aktivita (Po-210)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2072 (5000 - 5.0E+7) Bq g-1 2.6 %
  • Po-210 glass ampoule
115-MP-C007
Quantity: Hmotnostní aktivita (Pb-210)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2073 (5000 - 5.0E+7) Bq g-1 2.6 %
  • Pb-210 glass ampoule
115-MP-C007
Quantity: Hmotnostní aktivita (Ra-226)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2074 (40 - 4.0E+4) Bq g-1 1.2 %
  • Ra-226 glass ampoule
115-MP-C007
Quantity: Hmotnostní aktivita (Am-241)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2075 (5000 - 5.0E+7) Bq g-1 1 %
  • Am-241 glass ampoule
115-MP-C007
Quantity: Activity (Cr-51)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2076 (5.0E+6 - 1.0E+10) Bq g-1 7 %
  • Cr-51 radionuclide in 10 ml gel in the standard vial
115-MP-C007
Quantity: Activity (Ga-67)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2077 (5.0E+6 - 1.0E+10) Bq g-1 6 %
  • Ga-67 radionuclide in 10 ml gel in the standard vial
115-MP-C007
Quantity: Activity (Tc-99m)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2078 (5.0E+6 - 1.0E+10) Bq g-1 6 %
  • Tc-99m radionuclide in 10 ml gel in the standard vial
115-MP-C007
Quantity: Activity (In-111)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2079 (5.0E+6 - 1.0E+10) Bq g-1 7 %
  • In-111 radionuclide in 10 ml gel in the standard vial
115-MP-C007
Quantity: Activity (I-131)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2080 (5.0E+6 - 1.0E+10) Bq g-1 6 %
  • I-131 radionuclide in 10 ml gel in the standard vial
115-MP-C007
Quantity: Activity per unit volume (I-131)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2081 (0.001 - 1.0E+4) Bq cm-3 9 %
  • I-131 methyliodine or elementary iodium incorporated in the monitor
115-MP-C007
Quantity: Activity per unit volume (Kr-85)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2082 (0.1 - 200) Bq cm-3 9 %
  • Kr-85 calibrated amount of gas incorporated in the monitor
115-MP-C007
Quantity: Activity per unit volume (Xe-133)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2083 (1 - 1.0E+7) Bq cm-3 9 %
  • Xe-133 calibrated amount of gas incorporated in the monitor
115-MP-C007
Quantity: Activity (Am-241)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2084 (1000 - 1.0E+4) Bq 4 %
  • Am-241 marinelli beaker 450 ml or 1000 ml
115-MP-C007
Quantity: Activity (Th-232)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2085 (1000 - 1500) Bq 4 %
  • Th-232 marinelli beaker 450 ml or 1000 ml
115-MP-C007
Quantity: Activity (Ra-226)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2086 (1000 - 3000) Bq 4 %
  • Ra-226 marinelli beaker 450 ml or 1000 ml
115-MP-C007
EUR-RAD-CMI-2094 (4.0E+4 - 5.0E+8) Bq 2 %
  • Ra-226 Ra-226 in Pt-tube
115-MP-C007
Quantity: Activity (K-40)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2087 (1000 - 1500) Bq 4 %
  • K-40 marinelli beaker 450 ml or 1000 ml
115-MP-C007
Quantity: Activity (Cs-137)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2088 (1000 - 3000) Bq 4 %
  • Cs-137 marinelli beaker 450 ml or 1000 ml
115-MP-C007
Quantity: Activity (Co-60)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2089 (1000 - 3000) Bq 4 %
  • Co-60 marinelli beaker 450 ml or 1000 ml
115-MP-C007
EUR-RAD-CMI-2097 (7.0E+5 - 7.0E+5) Bq 1.6 %
  • Co-60 radionuclide between two polyethylene foils
115-MP-C007
Quantity: Activity (Pb-210)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2090 (1000 - 3000) Bq 4 %
  • Pb-210 marinelli beaker 450 ml or 1000 ml
115-MP-C007
Quantity: Activity (Eu-152)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2091 (1000 - 3000) Bq 4 %
  • Eu-152 marinelli beaker 450 ml or 1000 ml
115-MP-C007
EUR-RAD-CMI-2102 (6.0E+5 - 6.0E+5) Bq 2 %
  • Eu-152 radionuclide between two polyethylene foils
115-MP-C007
EUR-RAD-CMI-2104 (1.0E+4 - 1.0E+6) Bq 5 %
  • Eu-152 ICRP 23
115-MP-C007
Quantity: Activity (Rn-222)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2092 (2.0E+4 - 2.0E+6) Bq 3 %
  • Rn-222 stationary or flow mode
115-MP-C007
Quantity: Activity per unit volume (Rn-222)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2093 (0.3 - 2) Bq.cm-3 4 %
  • Rn-222 water solution Rn-222 without Ra-226
115-MP-C007
Quantity: Activity (Co-57)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2095 (5.0E+4 - 5.0E+4) Bq 2.4 %
  • Co-57 radionuclide between two polyethylene foils
115-MP-C007
Quantity: Activity (Mn-54)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2096 (4.5E+5 - 4.5E+5) Bq 2 %
  • Mn-54 radionuclide between two polyethylene foils
115-MP-C007
Quantity: Activity (Sr-85)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2098 (2.5E+5 - 2.5E+5) Bq 2.4 %
  • Sr-85 radionuclide between two polyethylene foils
115-MP-C007
Quantity: Activity (Y-88)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2099 (7.0E+5 - 7.0E+5) Bq 3 %
  • Y-88 radionuclide between two polyethylene foils
115-MP-C007
Quantity: Activity (Ce-139)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2100 (8.0E+4 - 8.0E+4) Bq 2.6 %
  • Ce-139 radionuclide between two polyethylene foils
115-MP-C007
Quantity: Activity (Ba-133)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2101 (2.5E+5 - 2.5E+5) Bq 2 %
  • Ba-133 radionuclide between two polyethylene foils
115-MP-C007
Quantity: Activity (Hg-203)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-2103 (1.5E+5 - 1.5E+5) Bq 2.8 %
  • Hg-203 radionuclide between two polyethylene foils
115-MP-C007
Quantity: Air kerma rate (Co-60)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-1001 (5.0E-5 - 10) Gy.h-1 3.5 %
  • Co-60 ISO-4037
115-MP-C006
EUR-RAD-CMI-1002 (5.0E-5 - 10) Gy.h-1 2 %
  • Co-60 ISO-4037
115-MP-C006
Quantity: Air kerma rate (Cs-137)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-1003 (3.0E-6 - 2) Gy.h-1 3.5 %
  • Cs-137 ISO-4037
115-MP-C006
EUR-RAD-CMI-1004 (3.0E-6 - 2) Gy.h-1 2 %
  • Cs-137 ISO-4037
115-MP-C006
Quantity: Air kerma rate (X-rays)
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-1005 (0.0002 - 0.02) Gy.h-1 4 %
  • X-rays, 50 kV to 420 kV ISO-4037 Narrow spectra N40-N300, 50 kV to 320 kV
115-MP-C006
EUR-RAD-CMI-1006 (0.0002 - 0.02) Gy.h-1 3 %
  • X-rays, 50 kV to 420 kV ISO-4037 Narrow spectra N40-N300, 50 kV to 320 kV
115-MP-C006
Quantity: Absorbed dose rate to water
Interní identifikátor Range Calibration and measurement capabilities Measurement conditions Procedures
EUR-RAD-CMI-1007 (0.2 - 10) Gy.h-1 3.5 %
  • Co-60 water phantom, 30 cm x 30 cm x 30 cm
115-MP-C006