You are here
View CMC
Field of measurement
- ionising radiation
Interní identifikátor
EUR-RAD-CMI-1002
Quantity
- Air kerma rate (Co-60)
Instrument or artifact
- Dosemeter
Calibration and measurement capabilities
Range | Calibration and measurement capabilities |
---|---|
5.0E-5 - 10 Gy.h-1 | 2 % |
Technical competence
- CIPM MRA
Guarantee of service
Guarantee for service | Department | Note |
---|---|---|
Ing. Vladimír Sochor | Unit of Primary Metrology of Ionizing Radiation | internal metrology service |
Measurement conditions
Method
Number of procedure | Title of procedure |
---|---|
115-MP-C006 | Determination of a reference value of kerma rate in air or absorbed dose rate in water in collimated photon beams for purposes of testing of measurement instruments. |